Commissioning Results of the SAPOTI Cryogenic Nanoprobe at the CARNAÚBA Beamline at Sirius/LNLS

28 Aug 2024, 11:20
20m
Saal B

Saal B

Invited talk 2. Beamline Innovations Mikrosymposium 2/3: Beamline Innovations

Speaker

Renan Ramalho Geraldes (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM)

Description

SAPOTI will be the second nanoprobe to be installed at the CARNAUBA (Coherent X-Ray Nanoprobe Beamline) beamline at the 4th-generation light source Sirius at the Brazilian Synchrotron Light Laboratory (LNLS) [1]. Working in the energy range from 2.05 to 15 keV, it has been designed for simultaneous multi-analytical X-ray techniques, including absorption, diffraction, spectroscopy, fluorescence and luminescence, and imaging in 2D and 3D. Highly-stable fully-coherent beam sizes between 30 and 120 nm, with monochromatic flux up to 1e11ph/s/100-mA/0.01%BW, are expected with an achromatic KB (Kirkpatrick-Baez) focusing optics, whereas a new in-vacuum high-dynamic cryogenic sample stage has been developed aiming at single-nanometer resolution images via high-performance 2D mapping and tomography [2].
A Loading Chamber (Figure 1) was designed to: i) preserve the vacuum level and cleanliness of the Main Chamber, with the KB system and a cryogenic nanopositioning sample stage, and ii) store up to six samples in cryogenic condition (around 100 K), using a Pulse Tube Cryocooler, in a carousel with an angular range of about 75°. The samples are preliminarily loaded in the Loading Chamber using a load-lock system (VCT500 Leica Microsystems) and a customized sample cartridge, which can carry up to three samples and gets engaged in the carousel. Then, for sample loading, a linear stage with a custom cryogenic gripper based on thermal inertial, gets a sample and goes toward the sample stage in the Main Chamber through an embedded gate valve, designed to separate the environments and with a 40mm diameter opening for the gripper tip. The key advantages with this cryogenic design are: i) the independence of complex liquid nitrogen circuits and infrastructure, ii) the avoidance of long conductive straps, which might be prone to fatigue and wear, including the release of particle debris on the sample and mirrors; and iii) compliance with high-throughput possibilities, given a loading-time target of a few tens of seconds, and robust and redundant automation. The overall alignment budget and the temperature-related limited amount of time required for this procedure were the biggest technical challenges for this project. This work presents the mechanical design, thermal models, alignment requirements, automation and operational procedures, including the assembly and offline commissioning results.
[1] Tolentino, H.C.N. et al., (2023). J. Electron Spectrosc. Relat. Phenom. 226 (147340).
[2] Geraldes, R.R. et al., (2023). AIP Conf. Proc. 2990 (040017).

I plan to submit also conference proceedings No

Primary authors

Francesco Lena (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Renan Ramalho Geraldes (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM)

Co-authors

Mr Erik Pereira (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Filipi Del Nero (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Gabriel Basilio (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Gabriel Brunheira (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) HÉLIO TOLENTINO (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Joao Vitor Bitteli França (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Lucas Perissinotto (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Matheus Henrique Siqueira da Silva (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Michel Machado (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Pedro Proença (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Rodrigo Gomes (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Vinicius Falchetto (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM) Mr Yago Marino (BRAZILIAN SYNCHROTRON LIGHT LAB - LNLS/CNPEM)