Characterization of the LCLS-II X-rays with Imagers, Power Meters and Fluorescence Intensity Monitors

29 Aug 2024, 17:45
15m
Saal F

Saal F

Contributed talk 1. Beamline Optics and Diagnostics Mikrosymposium 1/4: Beamline Optics and Diagnostics

Speaker

Philip Heimann (SLAC)

Description

The LCLS-II superconducting accelerator has begun operations and has produced x-rays since August 2024. The initial parameters are a photon energy range from 250 eV to 3.8 keV and a maximum repetition rate of 1 kHz. The absorption edges of solid filters were used to calibrate the photon energy and evaluate the FEL bandwidth from the hard x-ray undulator. The observed aluminum K absorption edge is shown in figure 1. A power meter measured the pulse energy at 3 keV. Imagers have determined the x-ray beam dimensions and the divergence. Power meters have ascertained the beamline transmission at the TMO and RIX instruments. The Fluorescence Intensity Monitor (FIM) provides a normalization of the intensity at the RIX endstations. The accuracy of the FIM has been improved by an analysis of the entire detector waveforms.

I plan to submit also conference proceedings Yes

Primary authors

Alberto Lutman (SLAC) Douglas Garratt (SLAC) Georgi Dakovski (SLAC) Heinz-Dieter Nuhn (SLAC) Kristjan Kunnus (SLAC) MIng-Fu Lin Philip Heimann (SLAC) Razib Obaid Stefan Moeller (SLAC) William Schlotter (SLAC)